Detector Image Not Found
Average User Rating:

Syft Technologies, Inc. - Voice200
Contact:

Syft Technologies, Inc.

1525 Park Manor Blvd, Suite 272

Pittsburgh, PA 15205

POC: Wesley Kenneweg

412-996-5915

System Cost: $225,000
Analysis Cost: $10-$20
Survey Source: Vendor and Internet Supplied Information
General Description
The Voice200 is a Selected Ion Flow Tube Mass Sprectrometer designed for both lab use and first responder use. It can measure TICs, TIMs, Explosives, CWAs, as whole air,breath and head space analysis down to parts per trillion ranges with little or no sample prep and in Real Time.
Technical Description
Voice200 takes ionized air through upstream quadrupole and filters all ions (except H3O+, NO+ and O2+) which are sequentially routed to flow tube every 20 milliseconds where the ions react with the analyte in a soft chemical first order reaction. The reaction products enter a downstream quadrupole where they are filtered again before being detected by a particle counter. The results are quantified in Real Time. The system can differentiate all isobars and some isomers, based on the use of the three precursor ions. Both SIM and Scan modes are possible and system can be operated by non-scientific operator.
Evaluation Criteria
THROUGHPUT:
  • 2 minutes or less for detection
  • Multiple samples, multiple tests/sample per run
  • System is continuous and provides real time analysis with no defined tests/samples
  • The system or device is currently fully automated
  • Device or system is intended for multiple detection assays
  • 0-1 solutions, buffer, eluents, and/or reagents
  • 2 components
  • 10-20 minutes is required for set-up
  • 1-2 steps are required for detection
LOGISTICS:
  • Very brief (minutes-hours) training and minimal technical skills
  • Approximately the size of a home dishwasher
  • More than 50 kg
  • Satellite, wireless and wired connections are available
  • System or device has 220V electrical requirement
OPERATIONS:
  • Can be used from 4°C to 37°C
  • Performance is not influenced by relative humidity
  • 5-10 years expected life
  • Results can be viewed in real-time
  • The system or device is currently fully autonomous
  • The system software is open but modification requires licensing
  • The system hardware is open but modification requires licensing
DETECTION:
  • Less than 250 µL
  • Superior specificity. System has a false alarm rate approaching zero (~0%)
  • 1x10-6 - 3x10-5 mg/m3
  • <1 ppb
  • Possible system could identify liquid chemical agent
Tier Selection
BioImageChemImageRadImage
Field Use
Mobile
Diagnostic
Analytical
Tier Key:
Tier 1 Top Tier
Tier 2 Second Tier
Tier 3 Third Tier
Tier 4 Fourth Tier
Tier 5 Bottom Tier
Scoring Analysis
Bar Chart Image Not Found
Impact Chart
Impact Chart Image Not Found
FIELD USE
MOBILE USE
DIAGNOSTIC USE
ANALYTIC USE