Detector Image Not Found
Average User Rating:

H3D Gamma - Polaris-H
Contact:

H3D Gamma

3250 Plymouth Rd Suite 203

Ann Arbor, MI 48105

Dr. Weiyi Wang

weiyi@h3dgamma.com

Tel: 734 661 6416 x111

www.h3Dgamma.com

System Cost: ~$100,000
Analysis Cost: N/A
Survey Source: Vendor Supplied Information
General Description
Polaris-H was developed for nuclear power plants as a complete solution for the identification, quantification, and localization of gamma-ray sources. It is a battery-operated system that weighs less than 10 pounds, with an integrated optical camera to overlay the gammaray image for each isotope the system identifies. The energy resolution is near 1% FWHM at 662 keV, and the detector can localize gamma-ray sources in all directions simultaneously. It is ready to use within two minutes of pressing the power button. It is airtight and watertight for easy decontamination. A tablet connects wirelessly for real-time display of the spectrum and image some distance away. The detector technology is 3D-position-sensitive CdZnTe (CZT).
Technical Description
The system is an imaging gamma spectrometer based on 3D-positionsensitive CdZnTe (CZT). This allow each gamma interaction to be located within the detector. This along with the energy measurement allows for the direction of the incoming photon to be calculated in real time. This information is displayed over a registered visual image to provide an energy selected view of the gamma emissions in the environment.
Evaluation Criteria
THROUGHPUT:
  • Between 2 and 15 minutes for detection
  • Multiple samples, multiple tests/ sample per run
  • System is continuous and provides real time analysis with no defined tests/samples
  • The system or device is currently semi-automated
  • Device or system is intended for multiple detection assays
  • 0-1 solutions, buffer, eluents, and/or reagents
  • 0 components
  • Less than 5 minutes is required for set-up
  • 1-2 steps are required for detection
LOGISTICS:
  • An afternoon of training and some technical skills required
  • Approximately the size of a toaster
  • Between 1 and 5 kg
  • Wireless and wired connections are available
  • System or device uses batteries
OPERATIONS:
  • Can be used from 4°C to 41°C
  • This system does not require consumable components
  • Performance is not influenced by relative humidity
  • Greater than 10 years expected life
  • Results can be viewed in real-time
  • The system or device is currently fully autonomous
  • The system software is closed and not available for modification
  • The system hardware is closed and not available for modification
DETECTION:
  • Not possible to achieve 510K clearance
  • Possible the system could receive FDA approval
  • This system does not test liquids and this question does not apply
  • Superior specificity. System has a false alarm rate approaching zero (~0%)
  • Total dose, dose rate and count rate with operator selection to show the display, may differentiate between types of radiation
  • Down to background level radiation (i.e., gamma 1 µR/hr)
  • Down to background level radiation, expressed in cpm or similar units
  • System is used for surveying
Tier Selection
BioImageChemImageRadImage
Field Use
Mobile
Diagnostic
Analytical
Tier Key:
Tier 1 Top Tier
Tier 2 Second Tier
Tier 3 Third Tier
Tier 4 Fourth Tier
Tier 5 Bottom Tier
Scoring Analysis
Bar Chart Image Not Found
Impact Chart
Impact Chart Image Not Found
FIELD USE
MOBILE USE
DIAGNOSTIC USE
ANALYTIC USE